Agilent Technologies 81134A Guide de l'utilisateur Page 47

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 92
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 46
Setting Up a Burst Signal Setting Up Generic and Advanced Signals
Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008 47
The instrument panel now looks as follows:
Channel Settings for the PRBS Signal To set the channel parameters:
1 Switch to the Channel page.
2 Set the signal mode to RZ.
3 In the Timing panel:
–Switch from Dcycle to Width and set the Width to 100 ps.
Set the Delay to 50 ps.
4 In the Levels panel, set:
The level format to Custom.
Amplitude to 2.0 V.
Offset to 0.0 V.
5 Enable the Channel 1 normal Output:
By pressing the respective softkey in the user interface:
– OR –
By pressing the normal Out softkey for channel 1 next to the
Output port.
Burst RepetitionInstrument