Resistance Measurement
Resistance measurement can be performed under either auto or manual test
conditions. Auto resistance measurement is performed in current source and
voltage measurement mode. The total auto resistance measurement error is
shown in the table below. The total error of a manual resistance measurement
can be calculated using the voltage and current accuracy information as shown
below.
Source I mode, manual ohm measurement (4-wire):
Total error = Vmeas/Isrc = R reading x (gain error % of V range + gain error
% of I range + offset error of I range / Isrc value %) + (offset error of V
range / Isrc value)
Source V mode, manual ohm measurement (4-wire):
Total error = Vsrc/Imeas = 1/ [1/R reading x (gain error % of I range + gain
error % of V range + offset error of V range / Vsrc value %) + (offset error of
I range / Vsrc value)]
Measurement speed = 1 PLC
Applicable for temperature: 23 °C ± 5 °C
Example of total error calculation:
I Source value = 1 mA at 1 mA range
V measure range = 2 V range
Total error (% reading + offset) = (0.02 % + 0.02 % + 200 nA/1 mA) + (350
μV/1 mA) = 0.06 % + 0.35 Ω
Typical Performance of Auto Resistance Measurement (4-wire), 2 V range
Range Resolution Test Current
Current
Range
Total Error
(% reading + offset)
2 Ω 1 μΩ 1 A 1 A 0.2 % + 0.00035 Ω
20 Ω 10 μΩ 100 mA 100 mA 0.06 % + 0.0035 Ω
200 Ω 100 μΩ 10 mA 10 mA 0.065 % + 0.035 Ω
2 kΩ 1 mΩ 1 mA 1 mA 0.06 % + 0.35 Ω
20 kΩ 10 mΩ 100 μA 100 μA 0.065 % + 3.5 Ω
200 kΩ 100 mΩ 10 μA 10 μA 0.06 % + 35 Ω
2 MΩ 1 Ω 1 μA 1 μA 0.095 % + 350 Ω
20 MΩ 10 Ω 100 nA 100 nA 0.18 % + 3.5 kΩ
200 MΩ 100 Ω 10 nA 100 nA 1.08 % + 35 kΩ
Measure
Speed
Measure To
Memory
Measure To
GPIB
Source Measure
To Memory
Source Measure
To GPIB
< 0.001 PLC 20000 12500 19500 12500
0.01 PLC 4500 3950 4500 3950
0.1 PLC 500 490 500 490
1 PLC 49 49 49 49
Operation reading rate varies by number of sweep steps. Number of sweep steps is specified from 1 to 2500.
System Speeds
Maximum Sweep Operation Reading Rates (reading/second) for 50 Hz