18
Program and Interface Capabilities
Programming Language
The Agilent B2961A/B2962A supports SCPI (Standard Commands for
Programmable Instruments).
Program Memory
Program memory allows you to store long strings of SCPI command lines
once into the B2961A/B2962A’s volatile memory and then recall those strings
multiple times while the program is executing using a single SCPI command. By
storing the command strings in memory, the time that would have been spent
sending those same commands over a communication bus is eliminated. For
tests that utilize lots of repeated code (such as subroutines), program memory
can dramatically reduce test times.
Maximum number of characters in program name:
32, consisting of alphabetic characters, numbers, hyphens and underscores.
Maximum memory size: 100 KB (2500 lines typical)
LXI
LXI Core Conformant. The Agilent B2961A/B2962A follows specified LAN
protocols and adheres to LXI requirements such as a built-in Web control server
and IVI-COM driver.
Ethernet: 10/100Base-T
USB2.0: USB-TMC488 protocol (Rear x 1)
GPIB: IEEE-488.2 compliant
USB file system
USB 2.0 high-speed mass storage (MSC) class device (Front x 1)
Digital I/O interface
Connector: 25-pin female D.
Input/output pins: 14 open drain I/O bits
Absolute maximum input voltage: 5.25 V
Absolute minimum input voltage: - 0.25 V
Logic low maximum input voltage: 0.8 V
Logic high minimum input voltage: 2.0 V
Maximum source current: 1 mA, Vout = 0 V
Maximum sink current: 50 mA, Vout = 5 V
5 V power supply pin:
Limited to 600 mA, solid state fuse protected.
Safety interlock pin:
One active high pin and one active low pin. Activation of both pin enables
output voltage > 42 V.
Maximum Number of Simultaneously Triggered Units (using Digital I/O)
1
: 8
Furnished accessories
Power cable
USB cable
Quick Reference (English)
Product Reference CD-ROM (including PDF manuals, Agilent B2900A Quick I/V
Measurement Software and drivers)
Agilent I/O Library Suite
Software and drivers
Agilent B2900A Quick I/V Measurement Software
The Agilent B2961A/B2962A includes PC-based Agilent B2900A Quick I/V
Measurement Software. This powerful software makes it easy to quickly setup
and perform I/V measurements and to display the measurement data in a table.