Agilent Technologies E1346A Manuel de service Page 21

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 66
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 20
Performance
Verification
Tests
The procedures in this section are used to test the multiplexer’s electrical
performance using the specifications in Appendix A - Specifications of the
Agilent E1346A Single Ended Relay Multiplexer User’s Manual as the
performance standard.
There are two performance verification tests for the relay multiplexers: Test
2-1: Closed-Channel Resistance Test and Test 2-2: DC Isolation Test.
These tests are suitable for incoming inspection, troubleshooting, and
preventive maintenance.
Wiring the
Test Fixture
A test fixture is required for the performance verification tests. Figure 2-1
shows typical connections using an Agilent E1346A terminal block for the
test fixture. You may want to order an extra terminal block to use as a test
fixture, so that you don’t have to re-wire each time the tests are performed.
The Agilent E1346A terminal block part number is E1346-80001.
Figure 2-1. Test Fixture Connections
Verification Tests 2-3
Vue de la page 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 65 66

Commentaires sur ces manuels

Pas de commentaire