Performance characteristics (continued)
Vertical system: scope channels
Scope channels MSO/DSO7xx2A: Ch 1 and 2 simultaneous acquisition
MSO/DSO7xx4A: Ch 1, 2, 3 and 4 simultaneous acquisition
Bandwidth (–3 dB)* MSO/DSO701xA: DC to 100 MHz
MSO/DSO703xA: DC to 350 MHz
MSO/DSO705xA: DC to 500 MHz
MSO/DSO710xA: DC to 1 GHz
AC coupled MSO/DSO701xA: 3.5 Hz to 100 MHz
MSO/DSO703xA: 3.5 Hz to 350 MHz
MSO/DSO705xA: 3.5 Hz to 500 MHz
MSO/DSO710xA: 3.5 Hz to 1 GHz
Calculated rise time (=0.35/bandwidth) MSO/DSO701xA: 3.5 nsec
MSO/DSO703xA: 1 nsec
MSO/DSO705xA: 700 psec
MSO/DSO710xA: 350 psec
Single-shot bandwidth MSO/DSO701xA: 100 MHz
MSO/DSO703xA: 350 MHz
MSO/DSO705xA: 500 MHz
MSO/DSO710xA: 1 GHz (in half-channel mode)
Range
1
MSO/DSO701xA, MSO/DSO703xA and MSO/DSO705xA: 2 mV/div to 5 V/div (1 MΩ or 50 Ω)
MSO/DSO710xA: 2 mV/div to 5 V/div (1 MΩ), 2 mV/div to 1 V/div (50 Ω)
Maximum input CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
Offset range ±5 V on ranges <10 mV/div; ±20 V on ranges 10 mV/div to 200 mV/div;
±75 V on ranges >200 mV/div
Dynamic range ±8 div
Input impedance 1 MΩ ± 1% || 14 pF or 50 Ω ± 1.5%, selectable
Coupling AC, DC
BW limit 25 MHz selectable
Channel-to-channel isolation DC to max bandwidth >40 dB
Standard probes 10073C shipped standard for each scope channel (1165A probes optional)
Probe ID Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
1 2 mV/div is a magnification of 4 mV/div setting for 350 MHz to 1 GHz models. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting
and 32 mV for 2 mV/div sensitivity setting.