Agilent Technologies N1022A Manuel d'utilisateur Page 32

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 42
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 31
32
The 10076A provides the
features you need to capture
fast, high-voltage signals. Its
compact design makes it easy to
probe today’s small, power
electronics components. Rugged
construction enables it to
withstand rough handling
without breaking.
High-Voltage Passive Probes
Agilent 10076A (100:1)
Measure voltages up to 4 kV peak
250 MHz bandwidth
Figure 6.3. Compact design and long
cable of the Agilent 10076A for probing
small, power electronics components.
Frequency
4000 V
1000 V
100 V
10 V
100 kHz
1 MHz
10 MHz
100 MHz
250 MHz
Volts dc + peak ac
Figure 6.4. Derating curve showing band-
width characteristics of the 10076A probe.
Specifications
Operating Characteristics
Bandwidth (–3 dB) 250 MHz
Rise time (Calculated) < 1.4 ns
Attenuation ratio 100:1
Input resistance 66.7 M (when terminated into 1 M)
Input capacitance Approximately 3 pF
Maximum input 4000 Vpk
Compensation range 6 - 20 pF
Probe readout Yes
Cable length 1.8 m
Ordering Information
Probe and Accessories
Part # Description Quantity
10076A High-voltage probe includes one retractable 1
hook tip, one ground bayonet, one IC probing tip,
one alligator ground lead, and a compensation
screwdriver.