Agilent Technologies N1022A Manuel d'utilisateur Page 33

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 42
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 32
33
Specifications
Operating Characteristics
Bandwidth (–3 dB) 50 MHz
Rise time (calculated) < 7 ns
Attenuation ratio 1000:1
Input resistance 100 M (when terminated into 1 M)
Input capacitance 1 pF
Compensation range 7 - 25 pF
Maximum voltage 15 kV dc, 10 kVrms ac, 30 kV dc + peak ac
Dimensions 2 cm (maximum width of probe stem after
handle) x 33 cm;
7.5 cm (maximum probe width at probe handle) x 33 cm
Ordering Information
Probe and Accessories
Part # Description Quantity
N2771A High-voltage probe includes one alligator ground 1
lead, one sharp probe tip, and one hook probe tip.
For safe and accurate insight into
very high-voltage designs check
out our model N2771A probe.
Typical applications include
PMTs, motor drives, high-voltage
High-Voltage Passive Probes
Agilent N2771A (1000:1)
Measure voltages up to 30 kV
dc + peak ac, 10 kVrms
50 MHz bandwidth
Superior protection and safety
Figure 6.5. Agilent N2771A for
measuring voltages up to
30 kV dc + peak ac, 10 kVrms.
switches, magnetrons, and mod-
ern projection systems. The
probe’s large size and rugged
construction provide superior
protection. The ground lead feeds
through the body of the probe
and protrudes behind the safety
barrier, keeping the ground
connection away from the
high voltage.
Frequency
100000 V
10000 V
1000 V
100 V
10 V
1 V
16.614 k
100 kHz
500 kHz
1 MHz
10 MHz
50 MHz
Volts dc + peak ac
Figure 6.6. Derating curve showing band-
width characteristics of the N2771A probe.