7-52 Service Guide N5230-90025
Repair and Replacement Procedures PNA Series Microwave Network Analyzers
Removing and Replacing the A41 Hard Disk Drive (HDD) N5230C
6. Navigate to D:\CalFiles\.
7. Copy the backup files prefixed with mxcalfile_ from the USB memory drive to the HDD.
8. After all files have been copied, remove the USB memory drive from the USB port.
9. Exit Windows Explorer.
Post-Repair Procedures
1. Adhere the new Certificate of Authenticity (license) label to the outer cover in the location specified in
Figure 7-1. on page 7-7.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal
procedure. Refer to Table 7-2 on page 7-58.