Agilent Technologies N5230C Manuel de service Page 41

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 296
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 40
Service Guide N5230-90025 3-9
PNA Series Microwave Network Analyzers Tests and Adjustments
N5230C Non-ANSI/NCSL Z540–1–1994 Verification
Non-ANSI/NCSL Z540–1–1994 Verification
For non-ANSI/NCSL Z540-1-1994, perform the preliminary checks and performance tests while stopping to
troubleshoot. Refer to Figure 3-2 for test flow. Troubleshoot and repair the first problem encountered without
continuing to other tests. After you troubleshoot, repair, and adjust, repeat the last failed portion and
generate a new set of data.
Figure 3-2 Non–ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart
Vue de la page 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 295 296

Commentaires sur ces manuels

Pas de commentaire