Agilent Technologies N5230C Manuel d'utilisateur Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 118
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
13
Table 5. 85052B Calibration Kit
N5230A/C- Option 220 (Standard Test Set and Standard Power Range)
Applies to the, N5230A/C Option 220 analyzers, 85052B (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port
calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Description Specification (dB)
45 MHz to
500 MHz
500 MHz to
2 GHz
2 to
8 GHz
8 to
20 GHz
Directivity 48 48 44 44
Source Match 40 40 33 31
Load Match 48 48 44 44
Reflection Tracking
±0.003
+0.02/°C
±0.003
+0.02/°C
±0.003
+0.03/°C
±0.006
+0.03/°C
Transmission Tracking
±0.010
+0.02/°C
±0.014
+0.02/°C
±0.062
+0.03/°C
±0.104
+0.03/°C
Vue de la page 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 117 118

Commentaires sur ces manuels

Pas de commentaire