Table 17. Test Port Input (Continued)
Description Specification Technical
Options 020, 120, 025, 125 Options 020, 120, 025, 125
Trace Noise Phase
6
(continued)
100 kHz IF bandwidth, ratioed measurement, nominal power at test port
300 kHz to
10 MHz
80 mdeg --
10 MHz to 6
GHz
30 mdeg --
6 GHz to
10.5 GHz
30 mdeg --
10.5 GHz to
13.5 GHz
60 mdeg --
600 kHz IF bandwidth, ratioed measurement, nominal power at test port
300 kHz to
10 MHz
-- 100 mdeg
10 MHz to 6
GHz
-- 60 mdeg
6 GHz to
10.5 GHz
-- 60 mdeg
10.5 GHz to
13.5 GHz
-- 80 mdeg
Table 17. Test Port Input (Continued)
Description Specification Typical
Options 020, 025, 120, 125, 220, 225, 420,
425, 520, 525
--
Reference Level Magnitude
Range +/-200 dB
Resolution 0.001dB
Reference Level Phase
Range +/-500°
Resolution 0.01°