54 User’s Guide
U3025AE10 Operational Check
8. To launch the Cal Set Viewer toolbar. Select Response > CAL > Manage CALS >
CAL Set Viewer.
Figure 52 Calibration, Cal Set Viewer
9. From the Cal Sets drop-down menu, select 999.1 and check Enable. Select the
Reflection Tracking(1,1) term in the center drop-down menu and ensure that the
Enable and Error Terms are selected.
NOTE You may also create a table on the network analyzer and enter the limit line
stimulus and response values. Select Limit Test ON and Limit Line ON, and
select Show Table. See Figure 53.
Figure 53 PNA-X Setting the Test Limits