User’s Guide 37
U3042AE12 System Setup with N5242A
6. Secure the network analyzer’s lower locking feet to the Test Set upper locking feet,
using the spring–loaded screws on the locking feet. Refer to Figure 28. If the network
analyzer's lower locking feet are not aligned with the screw holes in the Test Set's upper
locking feet, loosen the screws securing the feet to the instrument slightly to align.
Figure 28 Locking Feet Screws
NOTE There are two Lock-Feet kits available. Refer to “Contacting Agilent” on
page 94 for ordering information.
•PNA 5023-0132 (Kit includes locking feet and screws)
Test Set N5242-20138 (right foot) and N5242-20139 (left foot).
•Screw 0515-2317