User’s Guide 77
U3042AE12 Operational Check
8. To launch the Cal Set Viewer toolbar. Select Response > CAL > Manage CALS >
CAL Set Viewer.
Figure 61 Calibration, Cal Set Viewer
9. From the Cal Sets drop-down menu, select 999.1 and check Enable. Select the
Reflection Tracking(1,1) term in the center drop-down menu and ensure that the
Enable and Error Terms are selected.
NOTE You may also create a table on the PNA and enter the limit line stimulus and
response values. Select Limit Test ON and Limit Line ON, and press Show
Table. See Figure 62.
Figure 62 PNA-X Setting the Test Limits