Agilent Technologies N9340A Manuel d'utilisateur Page 142

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 172
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 141
6 SCPI Command Reference
132 N9340A User’s Guide
SWEep Subsection
Sweep Time
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
This command specifies the time in which the
instrument sweeps the display.
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
This command automatically selects the fastest
sweep time for the current settings.
Sweep Mode
[:SENSe]:SWEep:TIME:AUTO:MODE:
NORMal|FAST
[:SENSe]:SWEep:TIME:AUTO:MODE
This command toggles the sweep mode between
normal and fast.
*RST State: 1.076067 s
Range: 10 ms to 1000 s
Available unit: ks, s, ms, us
Key Access: [BW/SWP] > {Sweep Time}
*RST State: On
Key Access: [BW/SWP] > {Sweep Time}
*RST State: Normal
Key Access: [BW/SWP] > {Sweep} > {Sweep Mode}