Key Reference 5
N9340A User’s Guide 75
Log Pwr
Averages the data as appropriate for the
logarithmic scaled y axis.
When average type is set to Log Pwr, “LgAv
appears on the left side of the display.
Key access: [BW/SWP] > {Avg Type}
Pwr
Converts the scale from logarithmic to linear power
units, and then averages the power trace data.
With average type set to Pwr, “PAvg” appears on
the left side of the display.
Key access: [BW/SWP] > {Avg Type}
Voltage
Averages the data as appropriate for the linearly
scaled y- axis scale.
With average type set to Voltage, “VAvg” appears on
the left side of the display.
Key access: [BW/SWP] > {Avg Type}
There are actually four types of averaging
processes within an N9340A. All, except VBW
filtering, are affected by this setting. They are
listed as follows:
Trace averaging
Averages signal amplitudes on a trace- to- trace
basis. (Press Log Pwr or Pwr)
•Average detector
Averages signal amplitudes during the time or
frequency interval represents the result as a single
measurement point.
•Noise marker
Averages signal amplitudes across measurement
points to reduce variations for noisy signals.
VBW filtering
Performs a form of averaging of video signal.