Chapter 5 99
Setting Measurement Conditions
Selecting Sweep Direction
5. Setting Measurement
Conditions
Selecting Sweep Direction
DUT can be measured in the desired sweep direction when its characteristics has hysteresis
with the sweep parameter.
Step 1. Press the
[Sweep] key to display the sweep menu.
Step 2. Press the
DIRECTION [ ] key to select the desired sweep direction (pressing the key toggles
the direction).
NOTE Setting the sweep direction is applied commonly to traces A and B.
Softkey label Selected sweep direction
DIRECTION [UP] Upward (from sweep start value to sweep stop
value)
DIRECTION [DOWN] Downward (from sweep stop value to sweep
start value)