Chapter 3 Programming Your Counter for Remote Operation
Using Multiple Commands
3-20 Programming Guide
3
When multiple subsystem commands are sent in one program message,
the first command is always referenced to the root node. Subsequent
commands, separated by “;”, are referenced to the same level as the
preceding command if no “:” is present immediately after the command
separator (the semicolon).
For example, sending :SENS:AVER:COUN 5; STAT ON is equivalent
to sending:
:SENS:AVER:COUN 5
:SENS:AVER:STAT ON
or
:SENS:AVER:COUN 5;:SENS:AVER:STAT ON
The “:” must be present to distinguish another root level command.
For example:
:SENS:AVER:COUN 5;:INIT:CONT OFF
is equivalent to sending:
:SENS:AVER:COUN 5
:INIT:CONT OFF
If the “:”(which is following the “;” and is in front of INIT) is omitted, the
Counter assumes that the second command is “:SENS:AVER:INIT:CONT
OFF” and generates a syntax error.