Résumé du contenu

Page 1 - Agilent B2201A

Agilent TechnologiesAgilent B2200AFemto Leakage Switch MainframeAgilent B2201A14ch Low Leakage Switch MainframeUser’s Guide

Page 3

3-38 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusGPIB Displays the sub menu that provides the following functions.• EXECUTEDi

Page 4

4 Programming

Page 5

4-2 Agilent B2200 User’s Guide, Edition 4ProgrammingThis chapter describes the automatic control programming of the Agilent B2200.•“Programming Basics

Page 6 - Safety Summary

Agilent B2200 User’s Guide, Edition 4 4-3ProgrammingProgramming BasicsProgramming BasicsThis section explains the commands used for the fundamental sw

Page 7

4-4 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming BasicsFundamental CommandsThe following commands are used to set the fundamental switc

Page 8 - Safety Symbols

Agilent B2200 User’s Guide, Edition 4 4-5ProgrammingProgramming BasicsSwitch Control The following commands are used to control open/close of the spec

Page 9 - In This Manual

4-6 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming BasicsIn the channel_list, you can specify multiple channels by using comma (,) or col

Page 10

Agilent B2200 User’s Guide, Edition 4 4-7ProgrammingProgramming ExamplesProgramming Examples This section provides examples of control programs for th

Page 11 - Contents

4-8 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesConnecting Input-Output PathsThe following example connects instrument output

Page 12

Agilent B2200 User’s Guide, Edition 4 4-9ProgrammingProgramming ExamplesTable 4-3 Input-Output Connection ExampleImports Ivi.visa.interop '1Modul

Page 13

ContentsAgilent B2200 User’s Guide, Edition 41. Introduction Agilent B2200 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 14

4-10 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesUsing Bias Mode The following example uses the bias mode.Figure 4-3 Bias Mode

Page 15

Agilent B2200 User’s Guide, Edition 4 4-11ProgrammingProgramming ExamplesTable 4-4 Bias Mode ExampleImports Ivi.visa.interop '1Module Module1Sub

Page 16

4-12 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesUsing Ground Mode The following example uses the ground mode.Figure 4-4 Groun

Page 17

Agilent B2200 User’s Guide, Edition 4 4-13ProgrammingProgramming ExamplesTable 4-5 Ground Mode ExampleImports Ivi.visa.interop '1Module Module1Su

Page 18

4-14 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesUsing Couple ModeThe following example uses the couple mode.Figure 4-5 Couple

Page 19 - 1 Introduction

Agilent B2200 User’s Guide, Edition 4 4-15ProgrammingProgramming ExamplesTable 4-6 Couple Mode ExampleImports Ivi.visa.interop '1Module Module1Su

Page 20

4-16 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesSaving Input/Output LabelsYou can define labels to use for the input/output p

Page 21 - Agilent B2200 Series

Agilent B2200 User’s Guide, Edition 4 4-17ProgrammingProgramming ExamplesTable 4-7 Label Definition and Data Save ExampleImports Ivi.visa.interop &apo

Page 22 - Front Panel

4-18 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesDefining Comment for Internal MemoryYou can define comments for the internal

Page 23 - Figure 1-1 Front Panel View

Agilent B2200 User’s Guide, Edition 4 4-19ProgrammingProgramming ExamplesTable 4-8 Memory Comment Definition ExampleImports Ivi.visa.interop '1Mo

Page 24 - Rear Panel

Agilent B2200 User’s Guide, Edition 4ContentsSelf-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 25 - Figure 1-2 Rear Panel View

4-20 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationCapacitance CompensationWhen the capacitance/conductance measurement is p

Page 26 - Switch Modules

Agilent B2200 User’s Guide, Edition 4 4-21ProgrammingCapacitance CompensationRequired ConditionsThe following conditions must be satisfied to use the

Page 27

4-22 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationIn Figure 4-6, C2H, C2L, C3H, C3L are the compensation coefficients defin

Page 28 - Specifications

Agilent B2200 User’s Guide, Edition 4 4-23ProgrammingCapacitance CompensationTo Create Compensation Data FileThis section explains how to create the c

Page 29 - General Specifications

4-24 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationTable 4-9 Template Compensation Data FilesFile nameaa. <path>: driv

Page 30

Agilent B2200 User’s Guide, Edition 4 4-25ProgrammingCapacitance CompensationTable 4-10 Compensation Coefficients and Modifications2. Measure the R, L

Page 31

4-26 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationTo obtain compensation coefficientsObtain the compensation coefficients a

Page 32

Agilent B2200 User’s Guide, Edition 4 4-27ProgrammingCapacitance CompensationTo Perform Measurement and CompensationPerform the capacitance measuremen

Page 33 - Accessories and Options

4-28 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationTable 4-13 Capacitance Compensation Program Example#include "stdaf

Page 34

5 SCPI Command Reference

Page 35

ContentsAgilent B2200 User’s Guide, Edition 4To Set GPIB Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 36

5-2 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceThis chapter describes the following for Agilent B2200:• SCPI commands available to con

Page 37 - 2 Installation

Agilent B2200 User’s Guide, Edition 4 5-3SCPI Command ReferenceTextual Notation for Subsystem Commands CAPITAL LETTERS Capital letters are the minima

Page 38

5-4 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommands SummaryCommon Commands This section describes common commands and queries, whi

Page 39 - Requirements

Agilent B2200 User’s Guide, Edition 4 5-5SCPI Command Reference*CLS*CLS This command clears the Status Byte Register, the Standard Event Status Regist

Page 40 - Installation

5-6 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference*ESR?The following table shows the bits of the Standard Event Status Register and binar

Page 41 - Operating Environment

Agilent B2200 User’s Guide, Edition 4 5-7SCPI Command Reference*IDN?Semantics The following table shows the bits of the Standard Event Status Registe

Page 42 - Inspection

5-8 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference*OPC*OPC This command starts to monitor pending operations, and sets/clears the Operati

Page 43 - Installing the B2200

Agilent B2200 User’s Guide, Edition 4 5-9SCPI Command Reference*RST*RST This command performs an instrument reset. Status after *RST is shown below:Ch

Page 44 - To Connect the GPIB Cable

5-10 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference*SRE*SRE This command sets the Service Request “Enable” Register bits. 1 enables, 0 ma

Page 45 - To Install the Blank Panel

Agilent B2200 User’s Guide, Edition 4 5-11SCPI Command Reference*STB?The following is example for query:OUTPUT @Agb2200;"*SRE?" ENTER @Agb22

Page 46 - Self-Test

Agilent B2200 User’s Guide, Edition 4ContentsUsing Couple Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 47 - Output Connections

5-12 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference*TST?*TST? This query command executes an internal self-test, then returns the result.

Page 48 - Connector Plates

Agilent B2200 User’s Guide, Edition 4 5-13SCPI Command ReferenceCommand SummarySubsystem CommandsCommand SummaryROUT subsystem :ROUTe subsystem has co

Page 49

5-14 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommand Summary[:ROUT]:SYMB:CHAN card_number,channel,'string'[:ROUT]:SYMB:CH

Page 50

Agilent B2200 User’s Guide, Edition 4 5-15SCPI Command ReferenceCommand SummaryCommand Description Bias Mode Commands[:ROUT]:BIAS:CHAN:DIS:CARD card_n

Page 51 - To Install Interlock Circuit

5-16 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommand SummaryGround Mode Commands[:ROUT]:AGND:CHAN:DIS:CARD card_number Ground-disab

Page 52 - To Install LED Circuit

Agilent B2200 User’s Guide, Edition 4 5-17SCPI Command ReferenceCommand Summary[:ROUT]:AGND:UNUSED card_number,'enable_port'[:ROUT]:AGND:UNU

Page 53 - Anode(+)

5-18 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommand SummaryDIAG subsystem :DIAGnostic subsystem has commands for executing the sel

Page 54 - To Mount Connectors

Agilent B2200 User’s Guide, Edition 4 5-19SCPI Command ReferenceCommand SummarySYSTEM subsystem:SYSTem subsystem is a collection of functions that are

Page 55 - 14 11 11 14

5-20 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommand Summary:SYST:MEMO:COMM memory_number,'comment':SYST:MEMO:COMM? memor

Page 56 - Input Connections

Agilent B2200 User’s Guide, Edition 4 5-21SCPI Command Reference:DIAGnostic:TEST:CARD:CLEar:DIAGnostic:TEST:CARD:CLEar This command clears the relay t

Page 57

ContentsAgilent B2200 User’s Guide, Edition 4[:ROUTe]:AGND:CHANnel:DISable[:LIST] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24

Page 58

5-22 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:DIAGnostic:TEST:FRAMe:CLEarSyntax :DIAGnostic:TEST:CARD:STATe? card_numberQuery resp

Page 59 - Measurement Cable Length

Agilent B2200 User’s Guide, Edition 4 5-23SCPI Command Reference:DIAGnostic:TEST:FRAMe:STATe?Syntax :DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FP

Page 60

5-24 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:AGND:CHANnel:DISable:CARDExample OUTPUT @Agb2200;":DIAG:TEST:FRAM:STAT?

Page 61 - Maintenance

Agilent B2200 User’s Guide, Edition 4 5-25SCPI Command Reference[:ROUTe]:AGND:CHANnel:ENABle:CARDExample OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS (@

Page 62

5-26 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:AGND:PORTInput port is always the input Ground Port. So, the input ports in c

Page 63 - 3 Front Panel Operation

Agilent B2200 User’s Guide, Edition 4 5-27SCPI Command Reference[:ROUTe]:AGND[:STATe]Example OUTPUT @Agb2200;":ROUT:AGND:PORT 0,13"OUTPUT @

Page 64

5-28 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:AGND:UNUSEDExample OUTPUT @Agb2200;":ROUT:AGND:STAT 0,ON"OUTPUT @A

Page 65 - Operation

Agilent B2200 User’s Guide, Edition 4 5-29SCPI Command Reference[:ROUTe]:BIAS:CHANnel:DISable:CARD[:ROUTe]:BIAS:CHANnel:DISable:CARD This command bias

Page 66 - To Enable Light Pen

5-30 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:BIAS:CHANnel:ENABle:CARDExample OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB:C

Page 67 - To Control Switch Condition

Agilent B2200 User’s Guide, Edition 4 5-31SCPI Command Reference[:ROUTe]:BIAS:PORTInput port is always the input Bias Port. So, the input ports in cha

Page 68 - To load the setup

Agilent B2200 User’s Guide, Edition 4Contents:SYSTem:ERRor? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 69 - To Use Bias Mode

5-32 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:BIAS[:STATe]In this example, A will be 4.[:ROUTe]:BIAS[:STATe] NOTE You canno

Page 70 - To Use Ground Mode

Agilent B2200 User’s Guide, Edition 4 5-33SCPI Command Reference[:ROUTe]:CLOSe:CARD?[:ROUTe]:CLOSe:CARD? This query command returns channel_list of al

Page 71 - To Display Module Information

5-34 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:CONNection:RULEExample OUTPUT @Agb2200;":ROUT:CLOS (@10101,10202)"

Page 72 - To clear error buffer

Agilent B2200 User’s Guide, Edition 4 5-35SCPI Command Reference[:ROUTe]:COUPle:PORTThe connection sequence is used only for cards that have been set

Page 73 - LED matrix

5-36 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:COUPle:PORT:DETectAt *RST, no input ports are coupled.Syntax [:ROUTe]:COUPle

Page 74 - Switch Control Functions

Agilent B2200 User’s Guide, Edition 4 5-37SCPI Command Reference[:ROUTe]:COUPle[:STATe][:ROUTe]:COUPle[:STATe] This command controls the Couple Mode f

Page 75 - Channel Configuration Mode

5-38 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:FUNCtion[:ROUTe]:FUNCtion This command sets the channel configuration, which

Page 76 - Connection Rule

Agilent B2200 User’s Guide, Edition 4 5-39SCPI Command Reference[:ROUTe]:OPEN[:LIST][:ROUTe]:OPEN[:LIST]This command disconnects the input ports from

Page 77 - Connection Sequence

5-40 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:SYMBol:PORTSyntax [:ROUTe]:SYMBol:CHANnel card_number,channel_number,'s

Page 78 - Bias Mode

Agilent B2200 User’s Guide, Edition 4 5-41SCPI Command Reference:SYSTem:BEEP[:ROUTe]:SYMBol:PORT? port_numberQuery response symbol_string <newline

Page 79

ContentsAgilent B2200 User’s Guide, Edition 4agb220xa_cmdData_Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 80 - Ground Mode

5-42 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:SYSTem:CDEScription?Syntax :SYSTem:CCONfig? card_numberQuery response card_configur

Page 81

Agilent B2200 User’s Guide, Edition 4 5-43SCPI Command Reference:SYSTem:CPONFor B2201A:• In the Auto Config mode, "card_description" respons

Page 82 - Couple Mode

5-44 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:SYSTem:CTYPe?For the device reset state, see *RST.Syntax :SYSTem:CPON card_numberExa

Page 83

Agilent B2200 User’s Guide, Edition 4 5-45SCPI Command Reference:SYSTem:DISPlay:LEDSyntax :SYSTem:DISPlay:LCD stateEven though state is OFF, some sta

Page 84 - Display Functions

5-46 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:SYSTem:ERRor?Syntax :SYSTem:DISPlay:STRing 'string'Example OUTPUT @Agb220

Page 85 - GndLock Error GndBias

Agilent B2200 User’s Guide, Edition 4 5-47SCPI Command Reference:SYSTem:MEMOry:COMMentExample OUTPUT @Agb2200;":SYST:KLC ON":SYSTem:MEMOry:

Page 86 - Card Selection key

5-48 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:SYSTem:MEMOry:LOAD:SYSTem:MEMOry:LOADThis command restores the B2200 setup informatio

Page 87 - Function Edit

Agilent B2200 User’s Guide, Edition 4 5-49SCPI Command Reference:SYSTem:PEN:SYSTem:PEN This command enables/disables the light pen.Syntax :SYSTem:PEN

Page 88 - Function key group

5-50 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceStatus Reporting StructureStatus Reporting Structure This section describes the status

Page 89 - Settings: 01 [ ]

Agilent B2200 User’s Guide, Edition 4 5-51SCPI Command ReferenceStatus Reporting StructureFigure 5-1 Status Reporting Structure of B2200 The above fig

Page 90 - Channel 01: ENABLED

Agilent B2200 User’s Guide, Edition 4Contentsagb220xa_readStatusByte_Q. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 91

5-52 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceStatus Byte RegisterStatus Byte Register The B2200 Status Byte Register contains bits

Page 92 - Setup Menus

Agilent B2200 User’s Guide, Edition 4 5-53SCPI Command ReferenceStatus Byte RegisterIn general, use serial polling (not *STB?) inside interrupt servic

Page 93

5-54 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceService Request Enable RegisterService Request Enable Register The Service Request Ena

Page 94

Agilent B2200 User’s Guide, Edition 4 5-55SCPI Command ReferenceStandard Event Status RegisterStandard Event Status Register The Standard Event Status

Page 95

5-56 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceStandard Event Status Enable RegisterTable 5-2 Standard Event Status Register of B2200

Page 96 - Selftest Menu

Agilent B2200 User’s Guide, Edition 4 5-57SCPI Command ReferenceOutput QueueOutput Queue The Output Queue stores response messages until they are read

Page 97 - Exit key to return to the

5-58 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceOutput Queue

Page 98

6 VXIplug&play Driver

Page 99

6-2 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverThis chapter introduces the VXIplug&play driver available for Agilent B2200, and c

Page 100

Agilent B2200 User’s Guide, Edition 4 6-3VXIplug&play DriverSystem RequirementsSystem RequirementsThe following system environments are required.•

Page 102

6-4 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverInstalling VXIplug&play DriverInstalling VXIplug&play DriverThe installation f

Page 103 - :ROUT:CONN:SEQ ALL,BBM

Agilent B2200 User’s Guide, Edition 4 6-5VXIplug&play DriverDriver FunctionsDriver FunctionsTable 6-1 lists all the functions for the Agilent B220

Page 104 - Fundamental Commands

6-6 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsGround Mode agb220xa_groundPort Selects the input ground port of the s

Page 105 - Switch Control

Agilent B2200 User’s Guide, Edition 4 6-7VXIplug&play DriverDriver Functionsagb220xa_biasChanCardThis function bias-enables or disables all the ou

Page 106

6-8 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_biasChanList_QThis function will query the instrument for the

Page 107 - Programming Examples

Agilent B2200 User’s Guide, Edition 4 6-9VXIplug&play DriverDriver Functionsbport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card

Page 108 - Connecting Input-Output Paths

6-10 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_closeCard_QThis function will query the card for the channel

Page 109

Agilent B2200 User’s Guide, Edition 4 6-11VXIplug&play DriverDriver FunctionsThe parameter “closechan_list” is an array of integers with each inte

Page 110 - Using Bias Mode

6-12 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionscmd_str Instrument command (cannot exceed 256 bytes in length).size

Page 111

Agilent B2200 User’s Guide, Edition 4 6-13VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_cmdInt16_Q(ViSession vi, ViString c

Page 112 - Using Ground Mode

Notices© Agilent Technologies, Inc. 2004, 2005, 2009, 2011No part of this manual may be reproduced in any form or by any means (including electronic s

Page 113

1-2 Agilent B2200 User’s Guide, Edition 4IntroductionThis chapter describes the basic functions and features of the Agilent B2200A/B2201A (Agilent B22

Page 114 - Using Couple Mode

6-14 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_cmdRealThis function passes the cmd_str string to the instru

Page 115

Agilent B2200 User’s Guide, Edition 4 6-15VXIplug&play DriverDriver Functionscmd_str Instrument command (cannot exceed 256 bytes in length).resul

Page 116 - Saving Input/Output Labels

6-16 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_connRuleSeqThe function sets connection rule and connection

Page 117

Agilent B2200 User’s Guide, Edition 4 6-17VXIplug&play DriverDriver Functionscport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (car

Page 118

6-18 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsA device clear will abort the present operation and enable the instru

Page 119

Agilent B2200 User’s Guide, Edition 4 6-19VXIplug&play DriverDriver FunctionsInstrument errors may occur when you places the instrument in a bad s

Page 120 - Capacitance Compensation

6-20 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_esr_Q(ViSession vi, ViChar_VI_FAR e

Page 121 - Required Conditions

Agilent B2200 User’s Guide, Edition 4 6-21VXIplug&play DriverDriver Functions1 : sets ground disabled card.gnd_cardno Card number. 1 (card 1), 2

Page 122

6-22 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_groundChanList_Q(ViSession vi, ViIn

Page 123

Agilent B2200 User’s Guide, Edition 4 6-23VXIplug&play DriverDriver Functionsgnd_port Input port number to be set to the ground port. 1 to 14 (in

Page 124

Agilent B2200 User’s Guide, Edition 4 1-3IntroductionAgilent B2200 SeriesAgilent B2200 SeriesAgilent B2200 series is a switching matrix designed for s

Page 125

6-24 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsvi Instrument handle. This is VI_NULL if an error occurred during th

Page 126 - Triaxial Cable

Agilent B2200 User’s Guide, Edition 4 6-25VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_openList(ViSession vi, ViInt32_VI_F

Page 127

6-26 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_readStatusByte_Q(ViSession vi, ViPI

Page 128

Agilent B2200 User’s Guide, Edition 4 6-27VXIplug&play DriverDriver Functionsfile_name Compensation data file name. Use absolute path. If the val

Page 129 - 5 SCPI Command Reference

6-28 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_testExec_QThis function performs the diagnostics and returns

Page 130

Agilent B2200 User’s Guide, Edition 4 6-29VXIplug&play DriverDriver Functionsagb220xa_timeOut_QThis function returns the timeout value for driver

Page 131

6-30 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functions

Page 133

7-2 Agilent B2200 User’s Guide, Edition 4Error MessagesThis chapter lists and describes the error messages for Agilent B2200. An error message consist

Page 134

Agilent B2200 User’s Guide, Edition 4 7-3Error MessagesStandard SCPI Error MessagesStandard SCPI Error Messages Standard SCPI error messages have nega

Page 135

1-4 Agilent B2200 User’s Guide, Edition 4IntroductionFront PanelFront PanelThe Agilent B2200 series provides the front panel keys, the LCD, and the LE

Page 136

7-4 Agilent B2200 User’s Guide, Edition 4Error MessagesStandard SCPI Error Messages-103 Invalid separator An illegal character was received when a sep

Page 137

Agilent B2200 User’s Guide, Edition 4 7-5Error MessagesStandard SCPI Error Messages-120 Numeric data error An error was detected in a numeric paramete

Page 138

7-6 Agilent B2200 User’s Guide, Edition 4Error MessagesStandard SCPI Error Messages-148 Character data not allowed A character parameter is not allowe

Page 139

Agilent B2200 User’s Guide, Edition 4 7-7Error MessagesStandard SCPI Error MessagesExecution Error If syntax of a SCPI command header and parameter is

Page 140

7-8 Agilent B2200 User’s Guide, Edition 4Error MessagesStandard SCPI Error MessagesDevice-Dependent Errors -3XX errors indicate that an B2200 operatio

Page 141 - Subsystem Commands

Agilent B2200 User’s Guide, Edition 4 7-9Error MessagesStandard SCPI Error MessagesQuery Errors If the output queue control of the B2200 detects one o

Page 142 - Command Description

7-10 Agilent B2200 User’s Guide, Edition 4Error MessagesB2200 Specific Error MessagesB2200 Specific Error Messages These are the B2200-specific errors

Page 143 - Bias Mode Commands

Agilent B2200 User’s Guide, Edition 4 7-11Error MessagesB2200 Specific Error MessagesB2200 Channel Related Errors Error Number Error Message and Descr

Page 144 - Ground Mode Commands

7-12 Agilent B2200 User’s Guide, Edition 4Error MessagesB2200 Specific Error MessagesB2200 Card/Mode/Port Related Errors Error Number Error Message an

Page 145 - Couple Mode Commands

Agilent B2200 User’s Guide, Edition 4 7-13Error MessagesB2200 Specific Error Messages3017 Too many relays closed. Max 52 relays/card. Too many relays

Page 146

Agilent B2200 User’s Guide, Edition 4 1-5IntroductionFront PanelFigure 1-1 Front Panel View

Page 147

7-14 Agilent B2200 User’s Guide, Edition 4Error MessagesB2200 Specific Error Messages3026 Cannot use same port for Unused and CoupleThe couple mode an

Page 148

1-6 Agilent B2200 User’s Guide, Edition 4IntroductionRear PanelRear PanelThe Agilent B2200 series has four card slots for the switch modules, the GPIB

Page 149 - :DIAGnostic:TEST:CARD:STATe?

Agilent B2200 User’s Guide, Edition 4 1-7IntroductionRear PanelFigure 1-2 Rear Panel View

Page 150 - :DIAGnostic:TEST:FRAMe:CLEar

1-8 Agilent B2200 User’s Guide, Edition 4IntroductionSwitch ModulesSwitch ModulesThe Agilent B2200A and B2201A support the dedicated switch module, Ag

Page 151

Agilent B2200 User’s Guide, Edition 4 1-9IntroductionSwitch ModulesFigure 1-4 Switch Module Block Diagram OutputInput12345678910111213 CMH14 CML123456

Page 152

1-10 Agilent B2200 User’s Guide, Edition 4IntroductionSpecificationsSpecificationsThis section lists specifications and supplemental information for t

Page 153 - ENTER @Agb2200;A$

Agilent B2200 User’s Guide, Edition 4 1-11IntroductionSpecificationsGeneral Specifications Temperature range Operating: 5 °C to 35 °CStorage: -20 °C t

Page 154 - [:ROUTe]:AGND:PORT

DECLARATION OF CONFORMITY According to EN ISO/IEC 17050-1:2004Manufacturer’s Name: Agilent Technologies Singapore (International) Pte. Ltd. Manufactur

Page 155 - [:ROUTe]:AGND[:STATe]

1-12 Agilent B2200 User’s Guide, Edition 4IntroductionSpecificationsSwitch ModulesTable 1-1 lists the specifications when the switch module is install

Page 156 - [:ROUTe]:AGND:UNUSED

Agilent B2200 User’s Guide, Edition 4 1-13IntroductionSpecificationsSupplemental Characteristics for B2200A/B2210AOffset current 1 10 fA, I-V portIM

Page 157

1-14 Agilent B2200 User’s Guide, Edition 4IntroductionSpecificationsSupplemental Characteristics for B2201A/B2211A Offset current 1 100 fA, I-V portI

Page 158

Agilent B2200 User’s Guide, Edition 4 1-15IntroductionAccessories and OptionsAccessories and OptionsAgilent B2200 is furnished with the following acce

Page 159 - [:ROUTe]:BIAS:PORT

1-16 Agilent B2200 User’s Guide, Edition 4IntroductionAccessories and OptionsTable 1-2 Options and AccessoriesModel NumberOption Item Description B220

Page 160 - [:ROUTe]:BIAS[:STATe]

Agilent B2200 User’s Guide, Edition 4 1-17IntroductionAccessories and Options16494A Triaxial cable16494A-001 1.5 m length16494A-002 3 m length16494A-0

Page 161 - [:ROUTe]:CLOSe[:LIST]

1-18 Agilent B2200 User’s Guide, Edition 4IntroductionAccessories and Options

Page 163 - [:ROUTe]:COUPle:PORT

2-2 Agilent B2200 User’s Guide, Edition 4InstallationThis chapter describes requirements to install Agilent B2200 and the tasks for installation, and

Page 164 - [:ROUTe]:COUPle:PORT:DETect

Agilent B2200 User’s Guide, Edition 4 2-3InstallationRequirementsRequirements This section describes the following requirements for the Agilent B2200.

Page 165 - [:ROUTe]:COUPle[:STATe]

• HerstellerbescheinigungGERÄUSCHEMISSIONLpa < 70 dBam Arbeitsplatznormaler Betriebnach DIN 45635 T. 19• Manufacturer’s DeclarationACOUSTIC NOISE E

Page 166 - [:ROUTe]:OPEN:CARD

2-4 Agilent B2200 User’s Guide, Edition 4InstallationRequirements• Plug: BS 1363/A, 250 V, 10 A• PN: 8120-4420• Plug: AS 3112, 250 V, 10 A• PN: 8120-4

Page 167 - [:ROUTe]:SYMBol:CHANnel

Agilent B2200 User’s Guide, Edition 4 2-5InstallationRequirementsWARNING For protection from electrical shock, the power cable ground must not be defe

Page 168 - [:ROUTe]:SYMBol:PORT

2-6 Agilent B2200 User’s Guide, Edition 4InstallationInspectionInspectionCAUTION Before Opening Packing MaterialsThe Agilent B2200 contains the conden

Page 169 - :SYSTem:CCONfig?

Agilent B2200 User’s Guide, Edition 4 2-7InstallationInstalling the B2200Installing the B2200 This section describes how to install the Agilent B2200.

Page 170 - :SYSTem:CDEScription?

2-8 Agilent B2200 User’s Guide, Edition 4InstallationInstalling the B2200To Set the GPIB Address Every device on the GPIB bus must have a unique addre

Page 171 - :SYSTem:CPON

Agilent B2200 User’s Guide, Edition 4 2-9InstallationInstalling the B2200To Install the Switch ModuleWARNING To prevent electrical shock, turn off the

Page 172 - :SYSTem:DISPlay:LCD

2-10 Agilent B2200 User’s Guide, Edition 4InstallationSelf-TestSelf-TestNOTE To confirm the specificationsThe self-test and diagnostics checks the ope

Page 173 - :SYSTem:DISPlay:STRing

Agilent B2200 User’s Guide, Edition 4 2-11InstallationOutput ConnectionsOutput ConnectionsThis section describes how to connect the Agilent B2200 outp

Page 174 - :SYSTem:KLC

2-12 Agilent B2200 User’s Guide, Edition 4InstallationOutput ConnectionsConnector PlatesConnector plates (Table 2-1) are used for the connection betwe

Page 175 - :SYSTem:MEMOry:DELete

Agilent B2200 User’s Guide, Edition 4 2-13InstallationOutput ConnectionsFor Kelvin connection, use Kelvin triaxial cable listed in Table 2-2. To make

Page 176 - :SYSTem:MEMOry:SAVE

This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical

Page 177 - :SYSTem:VERSion?

2-14 Agilent B2200 User’s Guide, Edition 4InstallationOutput ConnectionsTable 2-3 Connection to the DUT InterfaceCAUTION Never connect the guard termi

Page 178 - Status Reporting Structure

Agilent B2200 User’s Guide, Edition 4 2-15InstallationOutput ConnectionsTo Make Interlock Circuit The interlock circuit is to prevent electric shock w

Page 179

2-16 Agilent B2200 User’s Guide, Edition 4InstallationOutput ConnectionsFigure 2-4 Dimensions of Interlock Switch (Agilent part number 3101-0302) Figu

Page 180 - Status Byte Register

Agilent B2200 User’s Guide, Edition 4 2-17InstallationOutput ConnectionsThe 4155/4156 semiconductor parameter analyzer's Intlk connector provides

Page 181

2-18 Agilent B2200 User’s Guide, Edition 4InstallationOutput ConnectionsNOTE To Check Interlock Circuit If you use the 4155/4156, you can easily check

Page 182

Agilent B2200 User’s Guide, Edition 4 2-19InstallationOutput ConnectionsTable 2-4 Recommended Parts Table 2-5 Dimensions of Connector Holes Descriptio

Page 183

2-20 Agilent B2200 User’s Guide, Edition 4InstallationInput ConnectionsInput ConnectionsThis section explains how to connect instruments to the Agilen

Page 184

Agilent B2200 User’s Guide, Edition 4 2-21InstallationInput ConnectionsTable 2-6 B2200 Input Connections Application Instrument B2200 Input Connector

Page 185 - Output Queue

2-22 Agilent B2200 User’s Guide, Edition 4InstallationInput ConnectionsIn the table above, • CMH/CML connectors are a pair of the AUX inputs.• The Agi

Page 186

Agilent B2200 User’s Guide, Edition 4 2-23InstallationMeasurement Cable LengthMeasurement Cable Length This section describes how to calculate the tot

Page 187 - 6 VXIplug&play Driver

Safety SummaryThe following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failur

Page 188

2-24 Agilent B2200 User’s Guide, Edition 4InstallationMeasurement Cable LengthTable 2-7 Guard Capacitances of B2200 Measurement Environment (Typical)

Page 189 - System Requirements

Agilent B2200 User’s Guide, Edition 4 2-25InstallationMaintenanceMaintenanceMaintenance should be performed periodically to keep the B2200 in good con

Page 190

2-26 Agilent B2200 User’s Guide, Edition 4InstallationMaintenance

Page 191 - Driver Functions

3 Front Panel Operation

Page 192

3-2 Agilent B2200 User’s Guide, Edition 4Front Panel OperationThis chapter explains the front panel operation and the switch control functions of the

Page 193

Agilent B2200 User’s Guide, Edition 4 3-3Front Panel OperationOperationOperationThis section describes operations of the Agilent B2200.• “To Initializ

Page 194

3-4 Agilent B2200 User’s Guide, Edition 4Front Panel OperationOperationTo Initialize Agilent B22001. Press the Shift key and the Local key. Then the L

Page 195

Agilent B2200 User’s Guide, Edition 4 3-5Front Panel OperationOperationTo Change Connection Rule1. Press the Rule key. The LCD displays the present se

Page 196

3-6 Agilent B2200 User’s Guide, Edition 4Front Panel OperationOperationTo Open All Switches1. Press the Open All key.2. Press the arrow key to display

Page 197

Agilent B2200 User’s Guide, Edition 4 3-7Front Panel OperationOperationTo Use Bias ModeBias mode cannot be set to ON when the ground mode is ON.1. Cha

Page 198

Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. R

Page 199

3-8 Agilent B2200 User’s Guide, Edition 4Front Panel OperationOperationTo Use Ground ModeGround mode cannot be set to ON when the bias mode is ON.1. C

Page 200

Agilent B2200 User’s Guide, Edition 4 3-9Front Panel OperationOperationTo Use Couple Mode1. Press the Couple Mode key to set the couple mode ON. Press

Page 201

3-10 Agilent B2200 User’s Guide, Edition 4Front Panel OperationOperationTo Read Error Message1. Press the Menu key. The setup menu will be displayed o

Page 202

Agilent B2200 User’s Guide, Edition 4 3-11Front Panel OperationOperationTo Set Remote Display ModeThis instruction enables or disables the data displa

Page 203

3-12 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsSwitch Control Functions This section introduces the switch con

Page 204

Agilent B2200 User’s Guide, Edition 4 3-13Front Panel OperationSwitch Control FunctionsChannel Configuration ModeThe Agilent B2200 provides the two ch

Page 205

3-14 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsConnection Rule Connection rule is used to specify if an input

Page 206

Agilent B2200 User’s Guide, Edition 4 3-15Front Panel OperationSwitch Control FunctionsConnection Sequence NOTE This is only for modules that are set

Page 207

3-16 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsBias Mode Bias mode is useful for connecting the same input to

Page 208

Agilent B2200 User’s Guide, Edition 4 3-17Front Panel OperationSwitch Control FunctionsNOTE Connection rule cannot be specified for the input bias por

Page 209

Safety SymbolsThe general definitions of safety symbols used on equipment or in manuals are listed below.Instruction manual symbol: the product will b

Page 210

3-18 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsGround ModeGround mode operation is similar to the bias mode op

Page 211

Agilent B2200 User’s Guide, Edition 4 3-19Front Panel OperationSwitch Control FunctionsNOTE Connection rule cannot be specified for the input ground p

Page 212

3-20 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsCouple Mode Couple mode is useful for making the Kelvin connect

Page 213

Agilent B2200 User’s Guide, Edition 4 3-21Front Panel OperationSwitch Control FunctionsNOTE Couple Port Detection FunctionThe Agilent B2200 provides t

Page 214

3-22 Agilent B2200 User’s Guide, Edition 4Front Panel OperationDisplay FunctionsDisplay FunctionsThe Agilent B2200 provides LED matrix, LCD, and 18 fr

Page 215

Agilent B2200 User’s Guide, Edition 4 3-23Front Panel OperationDisplay FunctionsLCDThe LCD displays the setup information and the status information a

Page 216 - VXIplug&play Driver

3-24 Agilent B2200 User’s Guide, Edition 4Front Panel OperationDisplay FunctionsTable 3-5 LCD Display ItemsNOTE In the remote mode and the default set

Page 217 - 7 Error Messages

Agilent B2200 User’s Guide, Edition 4 3-25Front Panel OperationFront Panel KeysFront Panel KeysThe Agilent B2200 provides LED matrix, LCD, and 18 fron

Page 218

3-26 Agilent B2200 User’s Guide, Edition 4Front Panel OperationFront Panel KeysFunction key groupThe Function key group is used to change the input po

Page 219 - Standard SCPI Error Messages

Agilent B2200 User’s Guide, Edition 4 3-27Front Panel OperationFront Panel KeysBias Mode Sets the bias mode ON or OFF. When the bias mode is set to ON

Page 220

In This Manual This manual is a user’s guide for Agilent B2200A/B2201A Switch Mainframe, and consists of the following chapters:• IntroductionDescribe

Page 221

3-28 Agilent B2200 User’s Guide, Edition 4Front Panel OperationFront Panel KeysGnd Ch (Shift+Open All)Displays the following menu.SET VIEW• SET displa

Page 222

Agilent B2200 User’s Guide, Edition 4 3-29Front Panel OperationFront Panel KeysEdit key groupThe Edit key group is used to select the switch module to

Page 223 - Execution Error

3-30 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusSetup MenusThis section explains the following setup menus displayed by pres

Page 224 - Device-Dependent Errors

Agilent B2200 User’s Guide, Edition 4 3-31Front Panel OperationSetup MenusFunction Tree The setup menu provides the following functions.• CONFIG• ADDR

Page 225 - Query Errors

3-32 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusCONFIG Displays the sub menu that provides the following functions.• ADDRESS

Page 226

Agilent B2200 User’s Guide, Edition 4 3-33Front Panel OperationSetup MenusERROR Displays the sub menu that provides the following functions.• DISPLAYD

Page 227 - B2200 Channel Related Errors

3-34 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusSelftest MenuPress the Shift key and the Menu key. The selftest menu is disp

Page 228

Agilent B2200 User’s Guide, Edition 4 3-35Front Panel OperationSetup MenusSELF_TEST Displays the sub menu that provides the following functions.• EXEC

Page 229

3-36 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusKEY Displays the sub menu that provides the following functions.• EXECUTEDis

Page 230

Agilent B2200 User’s Guide, Edition 4 3-37Front Panel OperationSetup MenusLED Displays the sub menu that provides the following functions.• EXECUTESta