Agilent B2200 User’s Guide, Edition 4 5-43
SCPI Command Reference
:SYSTem:CPON
For B2201A:
In the Auto Config mode, "card_description" response is one of following:
"B2211A 14x12 Low Leakage Switch Module"
"B2211A 14x24 Low Leakage Switch Module"
"B2211A 14x36 Low Leakage Switch Module"
"B2211A 14x48 Low Leakage Switch Module"
In the Normal Config mode:
"B2211A 14x12 Low Leakage Switch Module"
Example OUTPUT @Agb2200;":SYST:CDES? 1"
ENTER @Agb2200;A$
:SYSTem:CPON
This command resets the specified card. Status after this command is shown below:
Channel Configuration: Not changed.
Connection Rule: Free
Connection Sequence: Break Before Make
Bias Mode: Off
Bias Input Port: 10
Bias-enabled Channels: All channels (output ports) are bias-enabled.
Couple Mode: Off
Couple Input Port: Cleared.
Ground Mode: Off
Ground Input Port: 12
Ground-enabled Channels: Cleared.
Ground-enabled Ports: Cleared.
Card Channel Status: All relays are opened.
Self-Test Result: Not changed.
Input Port Symbol String: Not changed.
Output Port Symbol String: Cleared.