152 Chapter 11
Example Test Scenarios
Calibration Example
Example Test Scenarios
We then use Sequence Studio to add 5 analysis intervals to the acquisition, which appear on the graph as
5 blue bars. The bars can be positioned on the screen, both horizontally (to specify the timing and
duration of the interval of measurement) and vertically (to specify the expected power level during that
interval, so that there is a reference for pass/fail power results). For each interval, a separate combination
of measurements can be selected. In this case, Basic Tx Power is selected for each interval, as that is the
type of result needed for the calibration procedure.
Figure 11-4. Analysis Intervals Added to the Acquisition