Chapter 4 51
Coordinating the Sequencers
Include Source in Sequence is disabled
Coordinating the Sequencers
Once the analyzer list sequencer is initiated, the order of events is as follows:
The “dummy step” acquisition causes the analyzer list sequencer to trigger the source list sequencer.
The source list sequencer begins executing its sequence and generating an RF output to the DUT.
The first “real” acquisition (containing analysis steps) is triggered by the RF input from the DUT.