Agilent Technologies ENA Series Manuel d'instructions Page 180

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 251
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 179
180 Chapter 6
Replacement Procedure
A2 Receiver Board Replacement
A2 Receiver Board Replacement
Tools Required
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 10 in-lb); or 8mm with 9.5in-lbs can be
alternatively used (for reconnecting SMA connector)
Removal Procedure
Refer to Figure 6-10 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 166
Step 2. Remove the front panel as described in “Front Panel Removal” on page 167.
Step 3. Remove two TORX T15 screws (item 1).
Step 4. Disconnect the all semi-rigid cables (item 2).
NOTE The label on the semi-rigid cable shows the last two digits of the part number.
Step 5. Slide out the A2 receiver board (item 3).
Figure 6-10 A2 Receiver Removal
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
Vue de la page 179
1 2 ... 175 176 177 178 179 180 181 182 183 184 185 ... 250 251

Commentaires sur ces manuels