Introduction to the 34980A 1
Agilent 34980A Mainframe User’s Guide 17
Tota li z e r
The digital modules have 32- bit totalizer input channels for counting
pulses and measuring frequency. You can manually read the totalizer
count or you can configure a scan to read the count.
You can configure the totalizer to count on the rising edge or falling edge
of the input signal. The minimum rise/fall time for input signals to the
totalizer is 5
S.
The maximum count is 4,294,967,295 (2
32
- 1). The count rolls over to "0"
after reaching the maximum allowed value.
You can configure the totalizer to read without affecting the count or
reset the count to zero without losing any counts.
32 Bits
Count +
Count -
Gate
Gate
Totalizer