Features and Functions 3
Agilent 34980A Mainframe User’s Guide 91
Automatic Trigger Delays
If you do not specify a trigger delay, the instrument selects a delay
for you. The delay is determined by the function, range, integration time,
and ac filter setting as shown below.
DC Voltage, Thermocouple, DC Current (for all ranges):
Resistance, RTD, Thermistor (2- and 4-wire):
AC Voltage, AC Current (for all ranges):
Frequency, Period:
Digital Input, Totalize:
Integration Time Trigger Delay
PLC > 1
PLC 1
2.0 ms
1.0 ms
R a n g e
T r i g g e r D e l a y
( f o r P L C > 1 )
T r i g g e r D e l a y
( f o r P L C 1)
100
1 k
10 k
100 k
1 M
10 M
100 M
2.0 ms
2.0 ms
2.0 ms
25 ms
30 ms
200 ms
200 ms
1.0 ms
1.0 ms
1.0 ms
20 ms
25 ms
200 ms
200 ms
AC Filter Trigger Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
7.0 seconds
1.0 second
120 ms
AC Filter Trigger Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
600 ms
300 ms
100 ms
T r i g g e r D e l a y
0 seconds