Setting up an Analysis
Preparing the HPLC System
13 If both baselines are stable, set the Y-range for the VWD signal to 300 mAU.
14 Select the menu item RunControl -> Sample Info and enter information about
this application (Figure 28). Press OK to leave this screen.
Figure 28 Sample Information
15 Fill the content of an isocratic standard sample ampoule into a vial and seal
the vial with a cap.
16 Fill the injection syringe with the isocratic standard sample.
Enter:
Operator Name
Filename
Subdirectory
Sample Name
Comment