1-190 Chapter 1
Alphabetical Command Reference
PWMC
PWMC
Syntax
PWMC<EACS|OFF|ONES><num>; or PWMC<EACS|OFF|ONES>?;
Description
These commands select the type of power meter calibration desired, and set the drive port
cal power.
Front Panel Equivalents
Command Description Range Query Response
PWMCEACS Power meter calibration done with each
sweep.
100 to 100 dB
<0|1><
L
F
>
PWMCOFF Turns off power meter calibration. 100 to 100 dB
<0|1><
L
F
>
PWMCONES Power meter cal done on one sweep. A
calibration sweep should be taken (TAKCS)
after selecting a one sweep power meter
calibration, to ensure a valid calibration.
100 to 100 dB
<0|1><
L
F
>
Command Hardkey Softkey
PWMCEACS
PWMCOFF
PWMCONES
Cal
EACH SWEEP
Cal
PWR MTR CAL OFF
Cal
ONE SWEEP