Agilent Technologies 8753ET Guide de l'utilisateur Page 208

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 546
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 207
1-194 Chapter 1
Alphabetical Command Reference
RAWOFFS
RAWOFFS
Syntax
RAWOFFS<ON|OFF>; or RAWOFFS?;
Description
By turning raw offsets off (RAWOFFSOFF), a full two-port error correction can be performed
without including the effects of the offsets. It also saves substantial time at recalls and
during frequency changes (see SM8). Raw offsets follow the channel coupling. RAWOFFS and
SAMC are linked in that when one changes state, so does the other. See “Example 2G: Take4
— Error Correction Processed on an External PC” on page 7-48.
Front Panel Equivalents
Command Description Range Query Response
RAWOFFS Selects whether sampler and attenuator
offsets are on or off.
N/A
<0|1><
L
F
>
Command Hardkey Softkey
RAWOFFS
System
RAW OFFSET ON OFF