Agilent Technologies E6474A Guide de l'utilisateur Page 109

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 184
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 108
Get Assistance, if You Need It 6
Agilent E6474A User’s Guide 109
Testing the signal
1 In the Markers group, click Add and To Max. The level and
frequency are displayed at the top-left corner of the
Spectrum Analyzer display.
2 In the Band group, select Downlink.
3 In the Freq/Channel group, change the Center frequency
according to Table 4 on page 110.
4 On the signal generator, set the frequency according to
Table 4 on page 110.
You should see a signal trace showing a single spectral peak in
the middle of the display. Figure 52 on page 110 shows an
example trace.
Figure 51 Spectrum Display with Marker Set to a Center Frequency of 1960MHz(CDMA 1900 band system)
Vue de la page 108