Agilent Technologies Z5623A Option H48 5 -33
Performance Verification
Setting Up Limit Testing
Figure 5-36 The “Scale...” Button
Step 6. The Scale window shown in Figure 5-6 will pop up. When it does, set
the following values:
Scale Per Division = 20 dB
Reference Level = 0 dB
Reference Position = 9 Divisions
When these values have been entered, the Scale window will look like
Figure 5-37. Click the OK button.