Agilent Technologies Option H48 Multiport Test Set Z5623A Manuel d'utilisateur Page 11

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 16
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 10
11
Test sets (typical performance)
E5091A multiport test set
Designed to work with the ENA Series of network analyzers
Frequency range: 50 MHz to 8.5 GHz Switch type: Solid state
Connectors: Type-N (f) Switching speed: 3 ms
Impedance: 50 ohm I/O control: USB
Option Designed for: Ports Raw (uncorrected) Raw (uncorrected) Test set type
return loss
1
insertion loss
Port A,T2,R1+,R1-,R2+,R2-: Port A,T2,R1+,R1-,R2+,R2-:
23 dB, 50 to 300 MHz 3 dB, 50 to 300 MHz
25 dB, 300 MHz to 1.3 GHz 3 dB, 300 MHz to 1.3 GHz
19 dB, 1.3 to 3 GHz 4 dB, 1.3 to 3 GHz
12 dB, 3 to 6 GHz 5 dB, 3 to 6 GHz All paths except
007 ENA series 7 11 dB, 6 to 8.5 GHz 6 dB, 6 to 8.5 GHz between R1+ and R2+,
Port T1: Port T1: and R1- and R2-
18 dB, 50 to 300 MHz 5 dB, 50 to 300 MHz
20 dB, 300 MHz to 1.3 GHz 5 dB, 300 MHz to 1.3 GHz
16 dB, 1.3 to 3 GHz 7 dB, 1.3 to 3 GHz
12 dB, 3 to 6 GHz 8 dB, 3 to 6 GHz
9 dB, 6 to 8.5 GHz 9.5 dB, 6 to 8.5 GHz
Port A,T2,R1+,R1-,R3+,R3-: Port A,T2,R1+,R1-:
23 dB, 50 to 300 MHz 3 dB, 50 to 300 MHz
25 dB, 300 MHz to 1.3 GHz 3 dB, 300 MHz to 1.3 GHz
19 dB, 1.3 to 3 GHz 4 dB, 1.3 to 3 GHz
12 dB, 3 to 6 GHz 5 dB, 3 to 6 GHz All paths except
009 ENA series 9 11 dB, 6 to 8.5 GHz 6 dB, 6 to 8.5 GHz between R1+, R2+
Port T1,R2+,R2-: Port T1,R2+,R2-,R3+,R3-: and R3+, and R1-,
18 dB, 50 to 300 MHz 5 dB, 50 to 300 MHz R2- and R3-
20 dB, 300 MHz to 1.3 GHz 5 dB, 300 MHz to 1.3 GHz
16 dB, 1.3 to 3 GHz 7 dB, 1.3 to 3 GHz
12 dB, 3 to 6 GHz 8 dB, 3 to 6 GHz
9 dB, 6 to 8.5 GHz 9.5 dB, 6 to 8.5 GHz
1 Return loss of test set port not used in measurement path.