If the Out-of-Lock warning message appears,
verify that the voltage within the RF pulse “ON”
portion is zero volts (observed on the scope). If
phase-lock has been achieved, select Continue;
if phase-lock has not been achieved, select Re-try.
Phase lock indications include a flat zero-volt
trace (no presence of any beat note or search
signal).
7. When the measurement is complete, print the
results and save the measurement file.
Figure 3-2. Absolute phase noise measurement for an Agilent 83732B under pulsed
carrier conditions