Agilent Technologies E5500 Series Guide de l'utilisateur Page 57

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 87
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 56
57
Fragment Library Preparation Using the AB Library Builder
System: 5500 Series SOLiD
Systems User Guide
AppendixB Supplemental Procedures
Quantitate the DNA with the NanoDrop
®
ND-1000 Spectrophotometer
B
4.
Lift the sampling arm and load 2 µL of Nuclease-free Water onto the lower
measurement pedestal and lower the sampling arm:
5. In the dialog box, click OK and allow the instrument to initialize.
6. Lift the sampling arm and use Kimwipes
®
to remove water from the
measurement pedestal and the sampling arm.
7. Load 2 µL of the same buffer that was used to resuspend or elute the DNA onto
the measurement pedestal and lower the sampling arm.
8. Click Blank and allow the instrument to take a measurement:
9. Lift the sampling arm and wipe away the buffer from both the upper and lower
measurement pedestals with Kimwipes
®
. The instrument is now ready to take
readings.
10. Load 2 µL of DNA sample onto the lower measurement pedestal and lower the
sampling arm.
Vue de la page 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 86 87

Commentaires sur ces manuels

Pas de commentaire