20 Chapter 3
Operation Verification
Checking the Output Power
Perform the following procedure to check the signal generator output power:
1. Zero and calibrate the power sensor to the power meter as shown:
✔✔✔
Power Sensor,
Input: Type- N (m)
Agilent E9304A E- Series power sensor
or Agilent 8482A power sensor
Power Sensor,
Input: APC 3.5 (m)
Agilent E4413A E- Series power sensor
✔✔✔
Power Sensor,
Input: 2.4 mm (m)
Agilent 8487A Option H84 power sensor Option
H84 adds a Standards Lab Calibration
Adapter,
2.4 mm (f) to (f)
Agilent 11900B 2.4 mm (f) to (f) adapter
Adapter,
APC 3.5 (f) to (f)
Agilent 1250- 1749 APC 3.5 (f) to
(f) adapter
Adapter,
APC 3.5 (f) to
Type-N (m)
Agilent 1250- 1744
APC 3.5 (f) to Type- N (m) adapter
Table 3-4 Recommended Equipment for the E8267D
E8267D
Option
520
E8267D
Option
532
E8667D
Option
544
Test Equipment
Recommended Agilent
Test Equipment