Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide 115
Optimizing Performance
Using an Output Offset, Reference, or Multiplier
All frequencies entered are interpreted as being relative to this reference frequency.
Amplitude: Press Amptd > More > Amptd Ref Set
The amplitude displays 0.00 dB, indicating that this is the RF output amplitude “zero level.”
All amplitudes entered are interpreted as being relative to this reference amplitude.
Examples
To set a new frequency or amplitude reference, turn the frequency reference off, and then follow the
steps above.
Setting a Frequency Multiplier
Using a frequency multiplier, the signal generator can display a frequency that is the multiple
(positive or negative) of the output value.
Displayed Value = multiplier value × output frequency
Output Frequency = displayed value ÷ multiplier value
To set a frequency multiplier:
1. Press Frequency > Freq Multiplier > multiplier value > x.
2. Set the desired frequency.
The display equals the output frequency times the multiplier value.
Parameter
Example
#1
Example
#2
Example
#3
Comments
Reference: 50 MHz 50 MHz 2 GHz A reference value must be positive.
Entered (and displayed) Value: 2 MHz 2 MHz 1 GHz The entered value can be positive or negative.
Output Frequency: 52 MHz 48 MHz 1 GHz
The signal generator alerts you if the output frequency or
amplitude is out of range.