Agilent Technologies HP 3325B Manuel d'instructions Page 213

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 252
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 212
Chapter 6 Backdating
The Agilent 53131A Calibration Procedures
Assembly-Level Service Guide 6-11
6
CHANGE 1 (Firmware Revisions 3317, 3335, 3402, 3427) (Continued)
To Calibrate the Gain for Channels 1 and 2
1 Press any one of the arrow keys until CAL: GAIN 1 ? is displayed.
2 Press Enter key.
A scrolling message is displayed. Follow the instruction by performing the following
steps.
3 Connect a BNC tee connector to Channel 1 of the Counter.
4 Connect the Counter, dc power supply, and multimeter as shown in Figure 2-1.
Figure 2-1. Gain Calibration Setup
5 Adjust the power supply until the multimeter displays + 5.000 V ± 0.001 V.
6 Disconnect the digital multimeter from the Counter to prevent noise from being
introduced into the measurement.
7 Press Enter key.
The Counter momentarily displays CALIBRATING, and then it should display GAIN
1 PASS.
If the fail message is displayed, refer to the troubleshooting section in this chapter.
+
--
DC Power Supply
HP 53131A
Counter Digital Multimeter
BNC Tee
Vue de la page 212
1 2 ... 208 209 210 211 212 213 214 215 216 217 218 ... 251 252

Commentaires sur ces manuels

Pas de commentaire