Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
Assembly-Level Service Guide 1-61
1
2.8 to 3.0 GHz Sensitivity for Option 030 Only
1 Disconnect the Agilent 8663A Synthesized Signal Generator from the Counter
(see Figure 1-14), and connect the Agilent 8340B Sweep Oscillator to Channel 3
of the Counter.
2 Change the settings of the Sweep Oscillator to 3.0 GHz, 40 dBm.
3 Increase the power level in 1 dB steps until the Counter displays 3.0 GHz.
The Counter should display 3.0 GHz at 21.0 dBm.
4 Sweep the frequencies from 3.0 GHz to 2.8 GHz.
The Counter should read frequencies from 3.0 to 2.8 GHz at 21.0 dBm.
5 Record the actual reading in the Performance Test Record (Test 4, Line 3).
6 Disconnect the test setup.
200 MHz to 5.0 GHz Sensitivity for Option 050 only
1 Connect the 8340B Sweep Oscillator to Channel 2 of the counter.
2 Change the settings of the Sweep Oscillator to 5.0 GHz, 30 dBm.
3 Increase the power level in 1 dB steps until the Counter displays a stable reading
of approximately 5.0 GHz.
The Counter should display 5.0 GHz at 23 dBm.
NOTE Use a power meter and sensor to ensure the accuracy of the power being input.
4 Sweep the frequencies from 200 MHz to 5.0 GHz.
The Counter should read frequencies from 200 MHz to 5.0 GHz at 23 dBm.
5 Record the actual reading in the Performance Test Record (Test 4, Line 4).