4-14
Replacing Instrument Assemblies
To Remove the A2 Flat Panel Display
To Remove the A2 Flat Panel Display
1 Follow the instructions for removing the front panel on page 4-7.
2 Use a T-10 TORX driver to remove the four screws that secure the display to the front panel, as
shown in Figure 4-16.
Figure 4-16. Removing the Display Screws
3 Carefully lift the display out of the front panel.
Figure 4-17. Removing the Display