Hub Hi-Speed Tests 4
Notes on USB Electrical Compliance Testing 123
3 Click OK to close the Test Instructions dialog.
EL_40 Resume Timing Response
The USB automated test application will prompt you to perform these
steps:
1 On the Hub Test menu of the HS Electrical Test Tool, select RESUME
from the Hub Command drop down menu. Click [EXECUTE] once to
resume the hub from suspend. The captured transition should be as in
the figure below.