2-14
Making Mixer Measurements (Option 089 Only)
Conversion Loss Using the Frequency Offset Mode
Figure 2-11 Connections for R Channel and Source Calibration (IF Range)
Setting Measurement Parameters for the IF Range
1. From the front panel of the analyzer, set the desired receiver (IF) frequency and source
output power by pressing:
If the LO frequency is not set to 0 Hz, press:
2. To select the measurement trace, press:
The measurement trace is shown on the display.
3. Select the analyzer as the system controller:
Performing a Power Meter (Source) Calibration Over the IF Range
1. Calibrate and zero the power meter.
2. Set the power meter's address:
(where aa is the GPIB address of the power meter)
3. Select the appropriate power meter by pressing until the correct
model number is displayed (Agilent 436A or Agilent 438A/437).
System
INSTRUMENT MODE
FREQ OFFS MENU
Start 100 M/µ Stop 350 M/µ
LO FREQUENCY
0 x1
Meas
INPUT PORTS
R
Local
SYSTEM CONTROLLER
SET ADDRESSES
ADDRESS: P MTR/HPIB
aa
x1
POWER MTR [ ]