Solving Problems with Saving or Recalling Files
Solving Problems with Saving or Recalling Files
If you encounter a problem when you are storing files to disk, or the analyzer internal
memory, check the following list for possible causes:
Look in the analyzer display message area. The analyzer may show a message that will
identify the problem. Refer to the "Error Messages" chapter of the reference guide if you
view a message.
Make sure that you are NOT using a single-sided floppy disk in the analyzer disk drive.
Make sure that you are using a formatted disk.
Make sure that the disk has not been formatted with the LIF-HFS (hierarchical file
system) extensions as the analyzer does not support this format.
If You Are Using an External Disk Drive
Make sure that the analyzer is in system controller mode, by pressing
.
Make sure that you have connected the disk drive to ac power, switched on the power,
and connected an GPIB cable between the disk drive and the analyzer.
Make sure that the analyzer recognizes the disk drive's GPIB address, as explained in
"If You Are Plotting Measurement Results to a Disk Drive" on page 4-13.
Make sure that the analyzer recognizes the disk (drive) unit that you selected (0 or 1).
If the external disk is a hard disk, make sure that the disk volume number is set
correctly.
If the disk drive is an older HP 9122, it may not recognize the newer high density disks.
Substitute the GPIB cable.
Substitute the disk drive.