Page 228

REGIONAL SALES AND SUPPORT OFFICESFor more information about Agilent Technologies test and measurement products, applications, services, andfor a curr

Page 229

○C Copyright 2009 Agilent Technologies Agilent 4263B LCR Meter Operation Manual Manual Change Agilent Part No. N/A Jun 2009 Change 1 Add

Page 230

○C Copyright 2009 Agilent Technologies Change 4 Add TAR in 0 m Impedance Measurement Accuracy Test (Page 9-40 to 9-42) as follows. Impedanc

Page 231

○C Copyright 2009 Agilent Technologies 1 μF 100 k D ±0.0176 ± 0.00032 55.73 Note 1: Result = Measurement – Calibrated Value of Standard

Page 232

○C Copyright 2009 Agilent Technologies Measurement time: Med Test signal level: 500 mV Status PASS Standard Frequency [Hz] Param

Page 233

○C Copyright 2009 Agilent Technologies Change 5 Add TAR in 1 m Impedance Measurement Accuracy Test (Page 9-43) as follows. Measurement: Cap

Page 234

○C Copyright 2009 Agilent Technologies Change 6 Add TAR in 2 m Impedance Measurement Accuracy Test (Page 9-44) as follows. Measurement: Cap

Page 236

○C Copyright 2009 Agilent Technologies Change 7 Add TAR in 4 m Impedance Measurement Accuracy Test (Page 9-45) as follows. Measurement: Cap

Page 237

○C Copyright 2007 Agilent Technologies Agilent 4263B LCR Meter Operation Manual Manual Change Agilent Part No. N/A July 2008 Change 1 Cha

Page 238

○C Copyright 2007 Agilent Technologies Change 3 Change the Measurement Uncertainty value of DC Bias Level Accuracy Test (Page 9-39) to the

Page 239

○C Copyright 2007 Agilent Technologies Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test for 0 m Capacit

Page 240

○C Copyright 2007 Agilent Technologies Change 7 Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test for 0 m

Page 241

○C Copyright 2007 Agilent Technologies Change 11 Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test for 0

Page 242

○C Copyright 2007 Agilent Technologies Change 14 Change the Measurement Uncertainty value of 1 m Impedance Measurement Accuracy Test for 1

Page 243

○C Copyright 2007 Agilent Technologies Change 16 Change the Measurement Uncertainty value of 2 m Impedance Measurement Accuracy Test for 2

Page 244

○C Copyright 2007 Agilent Technologies Change 19 Change the Measurement Uncertainty value of 4 m Impedance Measurement Accuracy Test for 4

Page 245

Agilent 4263B LCR Meter Operation Manual Manual Change Agilent Part No. N/A July 2007 Change 1 Add the following steps to the step 2 of DC Bias Leve

Page 247

マニュアル チェンジ 変更 1 DC バイアス・レベル確度試験の手順 2(ページ 9-9)に以下の手順を追加して下さい。 2-a. Freq を押して測定信号周波数を 100 kHz に設定します。 2-b. Level を押して測定レベルを 50 mV に設定します。 変更 2 DC バ

Page 248

Agilent 4263B LCR Meter Operation Manual Manual Change Agilent Part No. N/A June 2007 Change 1 Add the following note to the “3-c. Turning On the 42

Page 249

マニュアル チェンジ 変更 1 電源投入と電源周波数の設定の手順 3-c(ページ 1-11)に以下の注記を追加して下さい。 注記 AC 電源周波数の設定値が一致していない場合、測定誤差が増大する可能性があります。必ず AC 電源周波数を正しく設定して下さい。 変更 2 測定確度の条件(ペー

Page 252

Safety SummaryWhen you notice any of the unusual conditions listed below, immediatelyterminate operation and disconnect the power cable.Contact your l

Page 256

4263BCautionDo not apply DC voltage or current to the UNKNOWN terminals. Donot apply External DC bias voltage more than +2.5V. Doing so willdamage the