536 Chapter7
Language Reference
MKSP Marker Delta to Span
MKSP Marker Delta to Span
Syntax
Figure 7-183 MKSP Syntax
Description
The MKSP command sets the frequency span equal to the frequency
difference between two markers on a trace. The start frequency is set
equal to the frequency of the left-most marker and the stop frequency is
set equal to the frequency of the right-most marker.
Example
10 INPUT "CONNECT THE 300 MHZ CALIBRATOR TO THE INPUT",Ans
$
20 OUTPUT 718;"IP;SNGLS;FA 270MHZ;FB 1275MHZ;TS;"
30 OUTPUT 718;"MKPK HI;MKD;MKPK NH;TS;MKSP;TS;"
40 END