616 Chapter7
Language Reference
ST Sweep Time
ST Sweep Time
Syntax
Figure 7-263 ST Syntax
Description
The ST command sets the sweep time. This is normally a coupled
function that is automatically set to the optimum value allowed by the
current instrument settings. Alternatively, you can specify the sweep
time. Note that when the specified sweep time is too fast for the current
instrument settings, the instrument is no longer calibrated and the
message MEAS UNCAL appears on the display.