ESG Family Signal Generators Hardkey and Softkey Reference
FM
User’s Guide 4-33
Trigger Out Polarity Neg Pos
This softkey toggles the polarity of the TTL signal that is output at the
rear panel TRIGGER OUT connector. When
Trigger Out Polarity Neg Pos is
set to
Pos, the output signal is asserted high (+5 V) at the start of the
swept-sine frequency modulation sweep and low (0 V) when the sweep
is concluded. Setting
Trigger Out Polarity Neg Pos to Neg reverse s the
polarity, where a high is 0 V and a low is +5 V.
This softkey is coupled to the
Trigger Out Polarity Neg Pos softkey in the
swept-sine AM,
Φ
M, and LF Out menus and the sweep/list menu. When
the polarity is toggled for this softkey in any one of these menus, it
affects all locations of the softkey.
Default Value: Pos
Softkey Location:
FM/ΦM > FM Waveform > Swept-Sine >
Trigger Out Polarity Neg Pos