Agilent Technologies 8133A Manuel de service Page 105

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 188
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 104
Troubleshooting 5
16800 Series Portable Logic Analyzers Service Guide 105
Adjust sampling positions using eye finder
1 Select the Thresholds and Sample Positions... button. The
Thresholds and Sample Positions dialog will appear.
2 In the “Buses/Signals” section of the Thresholds and
Sample Positions dialog, ensure that the check box next to
“My Bus 1” is checked.
3 Drag the blue bar for “My Bus 1” to approximately - 2 ns.
4 Select the plus sign to expand bus “My Bus 1”.
5 Click the Run button in the Thresholds and Sample
Positions dialog.
6 Ensure that an eye appears for each bit. Depending on
your test setup, the eye position may vary. Ensure that all
blue bars in the individual channel rows (excluding the
Vue de la page 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 187 188

Commentaires sur ces manuels

Pas de commentaire