Troubleshooting 5
16800 Series Portable Logic Analyzers Service Guide 95
Passing the wait test implies the module will operate
properly to user commands.
In case of error, the following diagnostic integer will be
displayed.
The four bit values have the following definition:
0 - passed
1 - failed to stop on break with no event wait
2 - failed on wait with setting of no event
3 - failed to stop on break or wait with wait any event
Instruction Interrupt Test
The instruction interrupt test verifies the WAIT IMB and
SIGNAL IMB instruction decoder. The module is configured
for a run, then verified that it is running. The module is
again configured for a run, then halted by a break command.
Passing the instruction interrupt test verifies that the
module will operate properly to user commands.
In case of error, the following diagnostic integer will be
displayed.
The four bit values have the following definition (errors are
or’ed):
0 - passed
1 - stopped without a break
2 - failed to stop from software
4 - failed to stop with break
8 - stopped by something other than break
Bit#: 15 - 12 11 - 8 7 - 4 3 - 0
Event D Event C Event B Event A
Bit#: 15 - 4 3 - 0
not used Status