138 16800 Series Portable Logic Analyzers Service Guide
6 Replacing Assemblies
To remove and replace a pattern generator card
Your product may consist of 1 pattern generator card and 1
measurement card.
1 Perform previous procedures:
“To prepare the instrument for disassembly" on
page 116.
2 Use a Torx T10 screwdriver to remove the 4 screws
securing the shroud to the frame.
3 Slide the shroud over the cables.
4 Use a gentle back- and-forth motion as you pull to
disconnect the cards from the connector on the module
interface board and then pull the cards out of the frame.
5 Replace the cards.
a Seat the lower card first. The pattern generator card
goes in the lower position.
b Press firmly to ensure the cards are fully seated in the
connector on the module interface board.
c Slide the shroud over the cables.
d Use a Torx T10 screwdriver and 4 screws to secure the
shroud to the frame.
T-10
screws
(4)
Shroud