Agilent Technologies 8133A Manuel de service Page 137

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 188
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 136
Replacing Assemblies 6
16800 Series Portable Logic Analyzers Service Guide 137
To remove and replace a measurement circuit board
1 Remove the 4 screws attaching the ground spring and
back panel to the circuit board, then remove the back
panel and the ground spring.
2 Replace the faulty circuit board with a new circuit board.
On the new board, ensure the 2x15 (30- pin) ribbon cable
is connected between J15 and J12.
3 Position the ground spring and back panel on the back
edge of the replacement circuit board. Install 4 screws to
connect the back panel and ground spring to the circuit
board.
4 Install the logic analyzer cables using the procedure “To
remove and replace measurement cables" on page 135.
5 Install the logic analyzer card using the procedure “To
remove and replace a measurement card" on page 134.
T-10
Screws
(4)
Ground
spring
Circuit
board
Back
panel
Vue de la page 136
1 2 ... 132 133 134 135 136 137 138 139 140 141 142 ... 187 188

Commentaires sur ces manuels

Pas de commentaire