Replacing Assemblies 6
16800 Series Portable Logic Analyzers Service Guide 135
If there is a pattern generator card in the frame, it
always goes in the lower position. If there are two logic
analyzer cards in the frame, the expander card goes in
the lower position.
a If your model has two logic analyzer (measurement)
cards, connect them as shown.
b Seat the lower card first.
c Press firmly to ensure the cards are fully seated in the
connector on the module interface board.
d Slide the shroud over the cables.
e Use a Torx T10 screwdriver and 4 screws to secure the
shroud to the frame.
To remove and replace measurement cables
1 Use a Torx T10 driver to remove the screws that secure
the measurement cable(s) to the back panel of the card.
2 Gently pull the measurement cable end connector(s) from
the circuit board connector(s).
3 If a measurement cable is faulty, replace the cable.
If you order a measurement cable, you also need to order
new labels. Probe cables shipped with the card are
already labeled. For part numbers and ordering
information, see “Measurement Card Replaceable Parts
List" on page 162.